3 Introduction

711 Liquino, Instructions for use

31


<SELECT>

<ENTER>




<SELECT>



<ENTER>





<QUIT>

dr i ve uni t : DOS1

The Dosino which is to be used can be selected here with <SELECT>

(DOSx refers to the connection number on the 711 Liquino). Accept
with <ENTER>.

Di spl ay:

>PI P dosi ng uni t

dos. uni t : def aul t

In this menu item the 711 Liquino is informed about the tubing of the

Dosino. Press <SELECT> until the dosing unit defined in the above
preparation step appears. (If the pipetting attachment is used this set-
ting can be left as ‘standard’). Confirm with <ENTER>.

Di spl ay:

par amet er s

>PI P r at es

As the ‘PIP dosing unit’ submenu contains no further menu items, an

automatic return is made to the submenu selection. All necessary pa-
rameters have now been entered! The parameter menu is exited with
<QUIT>.

Di spl ay:

PI P * * * * * * * *

r eady 0. 000 mL



If an inquiry is always to be made about the pipetting volume at the start of
the method then enter the value under ‘parameters/>PIP preselec-
tions/query pip. vol.’.

Various PIP methods can also be edited and then stored in the method
memory. Each of these methods can then be reloaded by pressing a few
keys. The method memory is described on page 128.


Carrying out the method

START

Z

When <START> is now pressed the display shows

Di spl ay:

Not pr epar ed yet !

Run PREP now?

ENTER

YES




<1> , <2>

Before the Dosino is used for the first time in the new configuration the

Dosino cylinder and the complete tubing system must be filled bubble-
free with reagent solution. Press <YES>.

Di spl ay:

wast e por t :

t i p( 1) or f l ask( 2) ?

Press <1> to empty the cylinder contents through the dosing tip, or

<2> to return them to the bottle. Take care that the tip of the dosing
tubing is located in a vessel. Wait until the 711 Liquino has ended the
preparation (PREP). The instrument then returns to its normal state.