4. Detailed description

711 Liquino, Instructions for use

89

4.5.2

SEQ mode parameters

PARAM

B

All settings in the parameter menu form a method and can be stored as
such.

Main menu:

par amet er s

>SEQ sequence

Open submenu with <ENTER>

par amet er s

>SEQ pr esel ect i ons

Use <> or <> to move down or up from one menu point to the
next

par amet er s

>SEQ r epor t s

Use <HOME> or <END> to jump to first or last menu point

Return to basic condition with <QUIT>

Submenus:

par amet er s

>SEQ sequence

Process sequence

Navigate with <ENTER>

>SEQ sequence

aut o st ar t 1

Number of times sequence is to be repeated
(confirm with <ENTER>)

1…99999,

infinite


(can be set with the <CLEAR> key)

>SEQ sequence st ep 1

command: NO OPERATI ON

Define step 1
(Select command with <SELECT>)

METHOD, PAUSE, TIMER,

SCAN, CONTROL, ENDSEQ,

NO OPERATION

Parameters of the individual commands can be
defined after confirmation with <ENTER>.

>SEQ sequence st ep 75

command: NO OPERATI ON

Further steps, up to step 75

METHOD, PAUSE, TIMER,

SCAN, CONTROL, ENDSEQ,

NO OPERATION.

Parameters of the individual commands can be
defined after confirmation with <ENTER>.


Up to 75 commands can be programmed in the sequence. Each
command forms a small submenu. After confirming the parameters of
the individual commands a new step is automatically attached to the
sequence.

In the sequence (step 1… x) quick navigation is possible by using the
cursor keys <> and <>. The <HOME> and <END> keys allow
a jump to the start or finish of the sequence.

With the <INSERT> key a new step can be inserted before the select-
ed command.
<DELETE> deletes the selected command from the sequence. The
following steps move up automatically.