4.4 XDOS mode - controlled dosing

711 Liquino, Instructions for use

78

>XDOS par amet er

f i l l r at e max mL/ mi n

Filling rate of dosing unit

max,

0.001…166.66

= highest possible filling rate (depends on type of dosing
unit.)

The filling rate is the adjustable piston stroke speed for aspirating the cylin-
der contents.

Reduce the filling rate for liquids with a higher density

or viscosity, as otherwise too rapid filling may cause an underpressure in

the cylinder.
The time for completely filling an empty dosing cylinder is approx. 20 sec-
onds at max. filling rate (including stopcock switching); at slower filling rates
it takes correspondingly longer.

>XDOS par amet er

aut o f i l l : on

Automatic filling after dosing

on, off

par amet er s

>XDOS dosi ng uni t s

Setting the dosing unit(s)

Navigate with <ENTER>

>XDOS dosi ng uni t s

set up: si ngl e

Use of one or two dosing drives

single,

tandem

= dosing with one dosing drive
= use of two dosing drives for
interruption-free dosing

>XDOS dosi ng uni t s

dosi ng dr i ve: DOS1

Dosing drive connection

DOS1, DOS2, DOS3, DOS4

>XDOS dosi ng uni t s

dos. uni t : def aul t

Name of dosing unit used

Selection list

see configuration '

>dosi ng uni t s'

>XDOS dosi ng uni t s

por t use: def aul t

Port allocations of dosing unit

default,

custom

= standard port occupancy, see page 76
= user-defined port allocation , see below

>XDOS dosi ng uni t s

dosi ng por t 1

Dosino port through which dosing is carried
out.

1…4

onl y f or ' cust om'


>XDOS dosi ng uni t s

f i l l i ng por t 2

Dosino port from which the dosing cylinder is
filled.

1…4

onl y f or ' cust om'


>XDOS dosi ng uni t s

wast e por t 4

Dosino port for excess reagent with PREP.

1…4

onl y f or ' cust om'


Maximal values for

dosing and filling rates

Cylinder vol. max. rate
2 mL

6.66 mL/min

5 mL

16.66 mL/min

10 mL

33.33 mL/min

20 mL

66.66 mL/min

50 mL 166.66 mL/min
These are pure piston
stroke rates; filling times
not taken into account.

For tandem dosing the
settings for two dosing

drives given alongside

must be carried out.

They are listed one

after another in the

menu.

It is a good idea to use

the same settings for

both dosing drives.