Yokogawa ADMAG AXF User Manual

Page 101

Advertising
background image

IM 1E10C1-E

11-9

11. PARAMETER LIST

Line size

0.01 to 300.0 mm/inch

100.0

W

Sets flow tube nominal size in selected unit.

b07

PV Damp

0.05 to 200.0 sec.

3.0

W

Sets damping time constant of output.

b02

Tag

Enter characters

W

Sets tag no. up to 8 characters.

Descriptor

Enter characters

W

Sets user-defined characters up to 16.

Message

Enter characters

W

Sets user-defined characters up to 32.

Date

01/01/00 to 12/31/99

0

W

Sets month/day/year.

Base bol unit

(00) k Cum

m

W

Selects volume unit of flow span.

b04

(01) Cum
(02) L
(03) Cucm
(04) M gal
(05) k gal
(06) gal
(07) m gal
(08) k bbl
(09) bbl
(10) m bbl
(11) u bbl
(12) m
(13) ft

Base time unit

(00) d

s

W

Selects time unit of flow span.

b05

(01) h
(02) min
(03) s

Velocity check

0 to 32.767 m/s

R

Displays span in m/s.

b13

Size unit

(00) mm

mm

W

Selects flow tube nominal size unit.

b06

(01) in

MF

0.2500 to 3.0000

1.0000

W

Sets meter factor.

b08

(Actual flow test data has been entered.)

Flow dir

(00) Forward

Forward

W

Selects flow direction.

b14

(01) Reverse

PV % rnge

–110.0 to 110.0%

R

Displays instantaneous flow rate in %.

A10

Rate limit

0 to 10%

5

W

Sets the level to reduce output fluctuation.

n03

Dead time

0 to 15 sec

0

W

Sets the dead time to reduce output fluctuation. When “0” is set,

n04

“Rate limit” is not available.

Power synch

(00) Yes

Yes

W

Selects synchronization between excitation frequency and

n05

(01) No

power frequency.

Power freq

47.00 to 63.00 Hz

50.00

W

Sets AC power frequency for DC power supply.

b12

Total unit

(00) n UNIT/P

PULSE/s

W

Selects totalization rate unit.

E01

(01) u UNIT/P

(“UNIT” in the selection on the left represents the “Base vol

(02) m UNIT/P

unit” selected.)

(03) UNIT/P
(04) k UNIT/P
(05) M UNIT/P
(06) PULSE/s

Total scale

0.0000 to 30000

0

W

Sets totalization rate.

E02

Total low cut

0 to 100%

3

W

Sets low cut width of totalization.

E03

Name

Data Range, Units

Default

Value

R/W

Description

Indicator

of SE

Advertising