Yokogawa ADMAG AXF User Manual

Page 65

Advertising
background image

IM 1E10C1-E

7-15

7. OPERATION VIA HART COMMUNICATOR

(3) Flow Units

For actual flow unit setting, the following parameters are provided.

<1> “PV Unit”

This parameter can be set only with HART Communicator. The selectable units are
shown below.

gal/min, Cum/h, L/min, m/s, ft/s

<2> Combination of “Base vol unit” and “Base time unit”

These parameters can be set with both ADMAG SE and HART Communicator. The
selectable units are shown below.

“Base vol unit”:

k Cum, Cum, L, Cucm, M gal, k gal, gal, m gal, k bbl, bbl,
m bbl, u bbl, m*, ft*
* “m” and “ft” can be combined only with /s.

“Base time unit”:

/d, /h, /min, /s

If the unit undefined in “PV Unit” is set in these parameters, the display of “PV
Unit” indicates “Spcl.”

(4) Meter Factor

The meter factor is engraved on the data plate of the combined flow tube.

1. Device setup

4. Detailed setup

1. Sensor

4. Sensor config

3. MF

ADMAG SE :
MF
1.0000
1.0000

DEL

DEL

ESC

ENTER

1

2

Enter “1.2345” with alphanumeric keys.

Press ENTER (F4).

Press SEND (F2) to set data to the ADMAG SE.

ADMAG SE :
Sensor config
1 Size unit in
2 Line size
3 MF 1.2345
4 Auto zero
5 Flow dir

HELP

SEND

HOME

ENTER

3

Call up “MF” setting display.

‘1.2345’

ADMAG SE :
MF
1.0000
1.2345

DEL

DEL

ESC

ENTER

Example: Set “1.2345” as meter factor.

3. MF

F4

(ENTER)

F2

(SEND)

Advertising