2 setting for test output using bt200, Setting for test output using bt200 -2 – Yokogawa ADMAG AXF User Manual

Page 84

Advertising
background image

IM 1E10C1-E

9-2

9. MAINTENANCE

(2) Status Output (For status output, alarm output or pulse output )

Call up “H01” (TEST MODE) on the setting mode and touch SET key to make the
data display.

Call up “H03” (STATUS OUT) and touch SET key.

Display

Description

Code

00
01

Normal operation
Test output

Content

Ч

2

Ч

2

SET

SET

SET

SET

Set “01” and touch SET key twice.

Select the output status and touch
SET key twice.

Code

00
01
02

Normal operation
Contact output status (ON)
Contact output status (OFF)

Content

9.1.2

Setting for Test Output Using BT200

(1) Current Output (Corresponding to flow rate, totalization display and pulse output value)

DATA

Display

Description

Select “TEST” in “H01: TEST MODE”.

Set the value in % of span in “H02:OUTPUT VALUE”. (The figure
shows a 100% setting.)
Flow rate, totalization display and pulse output functions correspond
to this set value.

PARAM
H01:TEST MODE
TEST
H02:OUTPUT VALUE
100 %
H03:STATUS OUT
NORMAL

DIAG

ESC

Code

NORMAL

TEST

Normal operation
Test output

Content

(2) Status Output (For status output, alarm output or pulse output status)

DATA

Display

Description

Call up “H01: TEST MODE” on the setting mode and select
TEST”.

Select the status to be output in “H03: STATUS OUT”.

PARAM
H01:TEST MODE
TEST
H02:OUTPUT VALUE
100 %
H03:STATUS OUT
NORMAL

DIAG

ESC

Code

NORMAL

TEST

Normal operation
Test output

Content

Code

NORMAL

CLOSED(ON)

OPEN(OFF)

Normal operation
Contact output status (ON)
Contact output status (OFF)

Content