6-8

6. OPERATION VIA BRAIN TERMINAL

Display

Description

Press ENTER key twice at an interval of a few seconds.
The display as shown on the left appears.

If the data is correct, press F4 key [OK].

The parameter panel is displayed again.

Select “B04:FLOW UNIT” by feeding the page with the arrow key.

Press ENTER key to display the data setup panel.

Select “m

3

” with the arrow key and press ENTER key twice at an

interval of a few seconds. Default:m
(If the security panel appears, enter the security code.*)

PARAM
B01: TAG NO

B02: DAMPING
3.0 S
B03: FLOW SPAN
30.0 m/s

DATA

DIAG

ESC

SET
B03:FLOW SPAN
1.00 m/s
30.0

DEL

CLR

ESC

SET
B03:FLOW SPAN
30.0 m/s

PARAM
B02: DAMPING
3.0 S
B03: FLOW SPAN
30.0 m/s
B04: FLOW UNIT
m

DIAG

DATA

ESC

SET
B04:FLOW UNIT
m
< km3 >
< m3 >
< l >
< cm3 >

ESC

ENTER

ENTER

×

2

NO

OK

F4

ENTER

×

2

SET
B04:FLOW UNIT
m3

NO

OK

F4

If the data is correct, press the F4 key [OK].

*km

: 10

3

X

m

cm

: (10

-2

X

m)

3

3

3

3

* For entry of the security code, see IM1C0A11-01E.