H) ᭹ span value setting – Yokogawa ADMAG AXF User Manual

Page 23

Advertising
background image

IM 1E10C1-E

5-4

5. FUNCTION AND DATA SETTINGS

(2) Span Settings by Infrared Switches on the Converter (Example: Flow Span 30.0m

3

/h)

Span Value Setting

Switch Operation

Display

Description

SET SHIFT INC

SET SHIFT INC

SET SHIFT INC

SET SHIFT INC

SET SHIFT INC

SET SHIFT INC

SET SHIFT INC

SET SHIFT INC

SET SHIFT INC

SET SHIFT INC

SET SHIFT INC

SET SHIFT INC

Change the display into the setting mode through the password
input mode. (Please refer to Chapter 4 “BASIC OPERATION
PROCEDURES.”)

By touching the SHIFT and INC key, select the parameter No.
b03”.

By touching the SET key, change the display into the data
changing mode.

By touching the INC key, change “1” into “0”.

By touching the SHIFT key, the cursor moves to the decimal
point.
The decimal point on the cursor can be moved by touching the
INC key.

And by touching the SHIFT key, blinking part moves to the
right.

By touching the INC key, change “0” into “3”.

By touching the SET key, whole Display is blinking.
And by touching the SET key again, the data is completed.

By touching the SET key again, the display goes back to the
number changing mode.

Defalt is set as 1.0000. Below is an example to change it into 0030.0.

(Setting is now completed)

(See 4.2.1 The initial
procedure to change the
display)

SET SHIFT INC

Advertising