3. INSTALLATION

3.

INSTALLATION

WARNING

This instrument must be installed by expert
engineer or skilled personnel. The procedures
described in this chapter are not permitted for
operators.

3.1 Installing Magnetic

Flow Converter

A signal cable (AM011) is used between the remote
type flow tube and the converter. The maximum signal
cable length is 300m (984ft).
The converter is mounted on a 2-inch (60.5mm outer
dia.) vertical or horizontal pipe. See Figure 3.1.1

Vertical Mounting

U-Bolt

F030101.EPS

2-inch pipe

Bracket

Nut

Horizontal pipe Mounting

2-inch pipe

U-Bolt

Bracket

Nut

Figure 3.1.1

Magnetic Flow Converter Installation

3.2 Wiring Precautions

This section is described wiring only for converter
side. Please see "Wiring" in SE***DJ/EJ Magnetic
Flow Tube User's Manual for flow tube side.

CAUTION

Confirm that all connections are corrected before
applying power to the instrument. Improper
wiring may damage the flow tube or converter.

NOTE

The terminal box cover is locked by special
screw. In case of opening the terminal box
cover, please use the Hexagonal Wrench
attached.

CAUTION

Be sure to lock the cover with the special screw
using the Hexagonal Wrench attached after
tightening the terminal box cover.

The external signal wirings are connected into the
terminal inside the converter. Please connect to each
terminal (Please refer to Figure 3.2.1) by taking off a
cover backside the converter.

Power supply

Protective ground

4 to 20 mA DC output

Pulse or alarm output

Dedicated Cable

AM011-4

Excitation Cable

A

B

C

EX2

EX1

Converter SE14

Flow Tube

F030201.EPS

Converter
Terminals

Flow Tube
SE***DJ/EJ

SA

A
B

SB

C

Ex1
Ex2

(See NOTE below)

A
B

(See NOTE below)

C

Ex1
Ex2

NOTE : Terminate those shielding wire
terminals using tape.

Figure 3.2.1 Wiring

3.2.1

Protective Grounding

CAUTION

Please be sure to connect protective grounding
of ADMAG SE with cable of 2mm

2

or larger

cross section in order to avoid the electrical
shock to the operators and maintenance engi-
neers and prevent the influence of external
noise. And further connect the grounding wire to
the

mark (100

or less).