8-1

8. ACTUAL OPERATON

8.

ACTUAL OPERATION

After you have installed the flow tube into the process piping, wired the input / output
terminals, set up the required parameters and performed the pre-operation zero adjust-
ment, the magnetic flowmeter should output an accurate flow signal from its terminals
as soon as the measured fluid begins to flow.
This section describes procedures of zero adjustment and alarms countermeasure.

8.1 Pre-Operation Zero Adjustment

This section describes three procedures for performing zero adjustment. Using the data
setting keys on the converter front panel, using a BT200 or using a Hart Communicator.
Zero adjustment is made to set the instrument output to 0% (4mA) when the flow rate is
0.

1. Zero adjustment should be done only when the fluid is filled in the flow tube and the

fluid velocity is completely zero by closing the valve.

2. Zero adjustment should be done prior to the other operation. For 20 seconds during

the zero adjustment, any setting cannot be accepted.

8.1.1

Zero Adjustment Using Data Setting Keys

Call up “C01” (ZERO TUNING) on the setting mode and touch SET key to make
the data display.

Call up “C02” (MAGFLOW ZERO) and touch SET key.

Touch SET key twice.

-ZEro-” is displayed for about 20 seconds.

It changes into the Flow Rate Display Mode after the zero adjustment is completed.

Display

Description

Code

00
01

INHIBIT
ENABLE

Content

Ч

2

Ч

2

SET

SET

SET

SET

Set “01” and touch SET key twice.

IMPORTANT