5.5.11 Totalization Switch Output

• In case the Internal Totalization Value increase over set level, an alarm signal is

output.

PLS/ALM+, PLS/ALM– terminals are for common use with pulse, alarm and other
status output functions. Therefore, in case this function is used, other functions are not
available to use.

Totalization switch can be selected as “05” in parameter
No. “b10”.

Switch level can be selected by calling up parameter No.
E06”.

When using reversed status (ON/OFF) is required,
it can be set in

“N02 Output Mode”.

Display

Description

Code

00
01
02
03
04
05

Pulse output
Alarm output
Forward / reverse flow measurement
Automatic two range switching
Alarm output at low flow limit
Totalization switch

Content

Totalization

999999

Setting value
10000

Status

Output

ON

Input

Default : 00

Default : 0