12. EXPLOSION PROTECTED TYPE INSTRUMENT

(3) Operation

WARNING

Note a warning label worded as follows.
Warning :

OPEN CIRCUIT BEFORE REMOV-
ING COVER.
OUVRIR LE CIRCUIT AVANT D’EN
LEVER LE COUVERCLE.

Take care not to generate mechanical spark
when access to the instrument and peripheral
devices in hazardous locations.

(4) Maintenance and Repair

WARNING

The instrument modification or parts replacement
by other than authorized representative of
Yokogawa Electric Corporation is prohibited and
will void CSA Explosionproof Certification.

12.4 SAA

(1) Technical Data

SA Certificate No.:

AUS EX 3764X

Type of Protection :

Ex d II C T6

Enclosure Type :

IP67

Excitation Circuit :

41V max. 6/6.25Hz

Temperature Code :

T6

Ambient Temperature : -20 to +60

°

C

(2) Installation

WARNING

• All wiring shall comply with local installation

requirements and local electrical code.

• In hazardous locations, the cable entry devices

shall be of a certified flameproof type, suitable
for the contidions of use and correctrly
installed.

(3) Operation

WARNING

• Open circuit before opening the covers.
• Take care not to generate mechanical spark

when access to the instrument and peripheral
devices in hazardous locations.

(4) Maintenance and Repair

WARNING

The instrument modification or parts replacement
by other than authorized representative of
Yokogawa Electric Corporation is prohibited and
will void the certification.