12 alarm output, 13 data settings enable / inhibit, 12 alarm output -17 – Yokogawa ADMAG AXF User Manual

Page 36: 13 data settings enable / inhibit -17

Advertising
background image

IM 1E10C1-E

5-17

5. FUNCTION AND DATA SETTINGS

5.5.12 Alarm Output

• This function is for status output from PLS/ALM+, PLS/ALM– terminals, when an

alarm occurs.

PLS/ALM+, PLS/ALM–terminals are for common use with pulse, alarm and other
status output functions.
Therefore, in case this function is used, other functions are not available to use.

• All of the alarms are able to output except empty pipe detection function that can be

selected in parameter No. “n07” as out of selection.

• The status goes from close to open (OFF) during alarming.

The alarm output can be selected “01” in parameter No.
b10” and P+, P– terminals are only for alarm output.

The empty pipe output selection can be set in parameter
No. “n07”.

Display

Description

Code

00
01
02
03
04
05

Pulse output
Alarm output
Forward / reverse flow measurement
Automatic two range switching
Alarm output at low flow limit
Totalization switch

Content

Content

Code

00
01

ALARM
NO ALARM

Default: 01

Default: 00

* Item “n” can be called up in setting “55” in parameter No. “L02”.

5.5.13 Data Settings Enable / Inhibit

• This function can inhibit to change all data except parameter No. “ L01”.

However, auto zero adjustment function can work, if it has been set in parameter No.
C01”.
And the preset totalization value function also can work, if it has been set in param-
eter No. “E04”.

The data settings inhibit item can be set “00” in
parameter No. “L01”.

Displayed

Description

Default: 01

Code

00
01

INHIBIT
ENABLE

Content

NOTE