Yokogawa ADMAG AXF User Manual

Page 44

Advertising
background image

IM 1E10C1-E

6-6

6. OPERATION VIA BRAIN TERMINAL

Display

Description

Press ENTER key twice at an interval of a few seconds.
The display as shown on the left appears.

If the data is correct, press F4 key [OK].

The parameter panel is displayed again.

Select “B07:NOMINAL SIZE” by feeding the page with the arrow
key.

Press ENTER key to display the data setup panel.

Enter “2.00” using the alphanumeric keys and press ENTER key
twice at an interval of a few seconds.
Default:10.00(for inch) 100.00(for mm)
(If the security panel appears, enter the security code.*)

PARAM
B04: FLOW UNIT
m
B05: TIME UNIT
/S
B06: SIZE UNIT
inch

DATA

DIAG

ESC

SET
B06:SIZE UNIT
mm
< mm

>

< inch

>

ESC

SET
B03:SIZE UNIT
inch

PARAM
B05: TIME UNIT
/S
B06: SIZE UNIT
inch
B07: NOMINAL SIZE
10.00 inch

DIAG

DATA

ESC

SET
B07:NOMINAL SIZE
10.00 inch
002.00

ESC

ENTER

ENTER

×

2

NO

OK

F4

ENTER

×

2

SET
B07:NOMINAL SIZE
2.00 inch

NO

OK

F4

If the data is correct, press the F4 key [OK].

“2.00”

PARAM
B05: TIME UNIT
/S
B06: SIZE UNIT
inch
B07: NOMINAL SIZE
2.00 inch

DIAG

DATA

ESC

The parameter panel appears again.
The setting is completed.

* For entry of the security code, see IM1C0A11-01E.