2 zero adjustment using bt200, Zero adjustment using bt200 -2 – Yokogawa ADMAG AXF User Manual

Page 78

Advertising
background image

IM 1E10C1-E

8-2

8. ACTUAL OPERATON

8.1.2

Zero Adjustment Using BT200

Display

Description

Call up the menu panel and select “C:ADJUST” with the arrow key.

Press ENTER key to call up the panel displaying the parameter
subitems.
Confirm the inverse video bar is on “C01:ZERO TUNING”.

Press ENTER key to call up the data setup panel and select
ENABLE”.

Press the ENTER key to set the data and return to the parameter sub
items panel. Select “C02:MAGFLOW ZERO”with the arrow key.

Pressing the ENTER key causes the data setup panel to be displayed.

While “C02:MAGFLOW ZERO” is displayed, press the ENTER
key twice at an interval of a few seconds.
Then, “AUTO ZERO” is now displayed.

After about 20 seconds, pressing the “F4” [OK] of the function key
causes the menu panel to be displayed. The zero correction is
displayed. Then the panel returns to data display mode.

MENU
A: DISPLAY
B: SET
C: ADJUST
D: DISP SEL
E: TOTAL SET
F: PULSE SET

HOME

SET

ADJ

ESC

ENTER

PARAM
C01: ZERO TUNING
ENABLE
C02: MAGFLOW ZERO
0.05
C60: SELF CHECK
GOOD

DATA

DIAG

ESC

PARAM
C01: ZERO TUNING
INHIBIT
C02: MAGFLOW ZERO
0.05
C60: SELF CHECK
GOOD

DATA

DIAG

ESC

SET
C01: ZERO TUNING
ENABLE

SET
C02: MAGFLOW ZERO
0.05
+00.05

CLR

ESC

SET
C02: MAGFLOW ZERO
AUTOZERO

FEED

NO

OK

PARAM
C01: ZERO TUNING
ENABLE
C02: MAGFLOW ZERO
0.02
C60: SELF CHECK
GOOD

DATA

DIAG

ESC

ENTER

ENTER

ENTER

ENTER

F4

Ч

2

Advertising