Self-diagnostics Using HART Communicator

Self-diagnostics of the ADMAG SE and check of incorrect data setting can be carried
out with the HART Communicator. There are two methods for self-diagnostics of the
ADMAG SE, self-diagnostics for every transmission and manually executing the “Self
test” command. When an error message appears, follow “8.2.3 Error Description and
Countermeasure.”

1. Device setup

2. Diag/Service

1. Test device

1. Test/Status

ADMAG SE :
Test/Status
1 Status
2 Self test

HELP

SAVE

HOME

ENTER

1

2

Select “Self test”.

If there is no error detected, “Self test OK” will be displayed.
Press OK (F4).
When an error occurs, an error message appears and the results
of self-diagnostics appear in the “Status”.

Call up “Status”.

The status menu is separated 3 groups. About items of each
group, see “8.2.3 Error Description and Countermeasure”.

Select the desired group.

If there is no error, the result of diagnostics is indicated as
OFF”. If “ON” is indicated, a countermeasure for that error is
necessary.

ADMAG SE :
Test/Status
1 Status
2 Self test

HELP

SAVE

HOME

OK

3

Call up “Test/Status” setting display.

ADMAG SE :
Self test OK

DEL

DEL

ABORT

OK

ADMAG SE :
Status
1 Status group 1
2 Status group 2
3 Status group 3

HELP

SAVE

HOME

OK

4

5

ADMAG SE :
group 1
Error OFF
uP fault OFF
EEPROM fault OFF
A/D fault OFF
Signal overflow OFF

HELP

DEL

ABORT

EXIT

F4

VWX

2

STU

1

(OK)