5.4 Power Frequency (For DC version only)

In case of DC power supply version, setting power frequency is required in areas where
the frequency differs. The meter is set to 50.00Hz at the factory.

Display

Description

Set the value in areas where the frequency differs in
b12”.

Default: 50.00Hz

5.5 Other Functions and Settings

5.5.1

Pulse Output

PLS/ALM+, PLS/ALM–terminals are for common use with pulse, alarm and other status
output functions. Therefore, in case this function is used, other functions are not
available to use.

(1) Pulse Output Overview

• By setting a pulse weight, a scaled pulse is transmitted to external counters or

measuring instruments.

Pulse Output Overview

Content

Item

Transistor contact output (contact capacity is 30V DC, 200mA)

PLS/ALM+, PLS/ALM–

When using these for pulse output, alarm output or status output are not
available as the terminals are used commonly.

Selection: DUTY50%, 0.5, 1, 20, 33, 50, 100ms

Min. 0.0001p/s, Max. 1000p/s

Output Specifications

Connecting
Terminals

Pulse Width

Output Rate

* Please refer to “3.2.6 Output Connection to External Instruments” for information how to connect external

instruments.

(2) Procedures for Setting Pulse Output

Example setting: 10 liter output per pulse in a flow rate span of

ᮀᮀᮀ m

3

/h

Display

Description

Pulse output can be set in parameter No. “b10” and
selected “00” (Pulse output).

Content

Code

00
01
02
03
04
05

Pulse output
Alarm output
Forward / reverse flow measurement
Automatic two range switching
Alarm output at low flow limit
Totalization switch

Default: 00

IMPORTANT

NOTE