8. ACTUAL OPERATON

8.2 Self-diagnostics Functions

᭿ The self-diagnostics function displays instrument internal errors, input/output signal

abnormalities, setting errors, and other problems.

᭿ When an alarm occurs, the alarm number that the error has occurred and normal data

are alternately indicated on the indicator of ADMAG SE.
However, alarms are only displayed during flow rate data display mode and when
parameter numbers are being changed in the setting mode. (Alarms are not displayed
when data items are being changed.)

᭿ The current output during the alarm occurrence can be selected from among the

followings. See “8.2.1 Output Status during Alarm Occurence.”
• 2.4mA or less
• 4mA
• HOLD
• 21.6mA or more

᭿ When the BT200 is used, alarms are displayed in “A60 to N60:SELF CHECK

parameter.

᭿ When the HART Communicator is used, alarms are displayed in “Status group 1 to

3” parameters. There are two methods for self-diagnostics, every transmission and
manually executing the “Self test” command. See “8.2.2 Self-diagnostics Using
HART Communicator.”

᭿ When an error message appears, follow “8.2.3 Error Description and Countermea-

sure.”.

8.2.1

Output Status during Alarm Occurrence

---

GOOD

Self test OK

ON

Normal

Normal

---

ERROR

Error

OFF

Fixed*

Stop

Err-02

uP FAULT

uP fault

OFF

21.6mA or more**

Stop

Err-03

EEPROM FAULT

EEPROM fault

OFF

21.6mA or more**

Stop

Err-05

A/D FAULT

A/D fault

OFF

Fixed*

Stop

Err-06

SIGNAL OVERFLOW

Signal overflow

OFF

Fixed*

Stop

Err-07

COIL OPEN

Coil open

OFF

Fixed*

Stop

Err-08

SPAN VEL.>10m/s

Vel. span>10m/s

OFF

Fixed*

Stop

Err-09

SPAN VEL.<0.3m/s

Vel. span<0.3m/s

OFF

Fixed*

Stop

Err-10

P.SPAN>1000p/s

P.span>1000p/s

OFF

Fixed*

Stop

Err-11

P.SPAN>500p/s

P.span>500p/s

OFF

Fixed*

Stop

Err-12

P.SPAN>25p/s

P.span>25p/s

OFF

Fixed*

Stop

Err-13

P.SPAN>15p/s

P.span>15p/s

OFF

Fixed*

Stop

Err-14

P.SPAN>10p/s

P.span>10p/s

OFF

Fixed*

Stop

Err-15

P.SPAN>5p/s

P.span>5p/s

OFF

Fixed*

Stop

Err-16

P.SPAN<.0001p/s

P.span<.0001p/s

OFF

Fixed*

Stop

Err-17

T.SPAN>1000p/s

T.span>1000p/s

OFF

Fixed*

Stop

Err-18

T.SPAN>.0001p/s

T.span<.0001p/s

OFF

Fixed*

Stop

Err-19

4-20 LMT ERROR

Analog lmt error

OFF

Fixed*

Stop

Err-20

EMPTY PIPE

Empty pipe

OFF

Fixed*

Stop

Err-21

MULTIRANGE ERROR

Multi range err

OFF

Fixed*

Stop

Err-22

COIL SHORT

Excitation failure

OFF

Fixed*

Stop

Err-23

EEPROM NO SET

EEPROM no set

OFF

21.6mA or more**

Stop

Err-24

EEPROM DEFAULT

EEPROM default

OFF

Fixed*

Stop

Err-25

Dev id not entered

OFF

Fixed*

Stop

Display detail

BRAIN Terminal

HART

Communicator

Indicator

of SE

Alarm

Contact

Output

Current Output

Total/Pulse

Output

Note) *Depending on the setting in “4-20 ALM OUT (Analog alarm out)”.

**When optional code /C1 is selected, current output becomes 2.4mA or less.