Select the “ scaled d/a trim ”. press ok (f4), Example 2: for the adjustment using a voltmeter – Yokogawa ADMAG AXF User Manual

Page 68

Advertising
background image

IM 1E10C1-E

7-18

7. OPERATION VIA HART COMMUNICATOR

ADMAG SE :
Trim analog output
1 D/A trim
2 Scaled D/A trim

HELP

SAVE

HOME

ENTER

1

2

Select the “Scaled D/A trim”.

Press OK (F4).

ADMAG SE :
WARN-Loop should be
removed from
automatic control

DEL

DEL

ABORT

OK

F4

VWX

2

(OK)

Example 2: For the adjustment using a voltmeter

ADMAG SE :
Fld dev output 4.000
mA equal to reference
meter?
1 Yes
2 No

HELP

SAVE

ABORT

OK

6

7

Ammeter reading: 4.000

Because the reading on the ammeter is 4.000 mA, select “Yes
and press ENTER (F4).
If the reading is not 4.000 mA, select “No”.
Repeat step until the ammeter reads 4.000 mA.

Press OK (F4), and the ADMAG SE outputs the output signal
of 100%.

Ammeter reading: 19.050

Carry out the same procedures as those described under .

Ammeter reading: 20.000

Returning fld dev to original output” appears.

Press OK (F4).

ADMAG SE :
Enter meter value
20.000

HELP

DEL

ABORT

ENTER

8

ADMAG SE :
Setting fld dev
output of 20mA

DEL

DEL

ABORT

OK

ADMAG SE :
Fld dev output 20.000
mA equal to reference
meter?
1 Yes
2 No

DEL

DEL

ABORT

ENTER

9

10

ADMAG SE :
NOTE-Loop may be
returned to automatic
control

HELP

DEL

ABORT

OK

F4

(OK)

F4

(ENTER)

‘19.050’

F4

(ENTER)

F4

(ENTER)

F4

(OK)

5

5