4-3

4. BASIC OPERATING PROCEDURES

4.2.1

The Initial Procedure to Change the Display Mode

• The procedure of changing the display from the Flow Rate Display Mode into the

Setting Mode by using infrared switches on the converter is described as follows.

(1) Procedure in General

The Flow Rate

Display Mode

The Pass-word

Input Mode

The Setting

Mode

SET key

(for 3 sec.)

Password Input

(It is always “4321”.)

(2) Procedure in Detail

Switch Operation

Display

Description

SET SHIFT INC

SET SHIFT INC

SET SHIFT INC

SET SHIFT INC

SET SHIFT INC

SET SHIFT INC

SET SHIFT INC

To go to the Setting Mode, it is needed to go to the Password
Input Mode first. Please touch the SET key for 3 sec., and the
display goes into the Password Input Mode.

By touching the SET key again, the display goes into the
Number Input Mode. Please input the Password as follows, by
touching the SHIFT and INC keys.

By touching the INC key for some moment, change “0” into
4”.

By touching the SHIFT key, the cursor moves to the next digit.

By touching the INC key for some moment again, change “0
into “3”.

By continuing this for the rest of two digits, please change the
whole digits into “4321”.

By touching the SET key, whole Display is blinking.
And by touching the SET key again, setting the password is
completed.

By touching the SET key again, the display goes to the number
changing mode.

(Setting is now completed)

SET SHIFT INC

(example)

(for 3 sec.)

(After power on)

In the Password Input Mode, if keys were not operated for more than 20 sec., or if
correct password were not set, the display goes back into the Flow Rate Display Mode
automatically.