IM 1E10C1-E

11-3

11. PARAMETER LIST

Indicator

Name

R/W

Description

Data Range, Units

BRAIN

Terminal

Decimal

Point

Default

Value

B00

SET

R

B01

TAG NO

ASC II 16 characters

W

b02

B02

DAMPING

0.5 to 200 sec

1

3.0

W

b03

B03

FLOW SPAN

1 to 3.0000

0 to 4

1.0000

W

b04

B04

FLOW UNIT

(00) km

3

m

(01) m

3

(02) l
(03) cm

3

(04) M gal
(05) k gal
(06) gal
(07) m gal
(08) k bbl
(09) bbl
(10) m bbl
(11)

µ

bbl

(12) m
(13) ft

b05

B05

TIME UNIT

(00) / d

/ s

W

(01) / h
(02) / m
(03) / s

b06

B06

SIZE UNIT

(00) mm

mm

W

(01) inch

b07

B07

NOMINAL SIZE

1 to 3000.0

1

100.0

W

b08

B08

MF

0.2500 to 3.0000

1.000

W

b10

B10

OUTPUT FUNC

(00) PULSE OUT

PULSE

W

(01) ALARM OUT

OUT

(02) BI DIRECTION
(03) AUTO 2 RANGES
(04) LOW ALARM
(05) TOTAL SWITCH

b11

B11

4 - 20 ALM OUT (00) 2.4 mA OR LESS

21.6 mA

W

(01) 4.0 mA

OR

(02) HOLD

MORE

(03) 21.6 mA OR MORE

b12

B12

POWER FREQ

47.00 to 63.00 Hz

2

50.00

W

Major parameter setting.

Sets tag no. up to 16 characters.

Sets damping time constant of output.

Sets flow span in selected unit.

Selects volume unit of flow span.

Selects time unit of flow span.

Selects flow tube nominal size unit.

Sets flow tube nominal size in selected unit.

Sets meter factor.

Selects status output functions.

Selects current output during alarm
occurrence.

Sets AC power frequency for DC power

(10

-2

X

m)

3

(liter)

(10

3

X

m

3

)