3 zero adjustment using hart communicator, Zero adjustment using hart communicator -3 – Yokogawa ADMAG AXF User Manual

Page 79

Advertising
background image

IM 1E10C1-E

8-3

8. ACTUAL OPERATON

8.1.3

Zero Adjustment Using HART Communicator

1. Device setup

2. Diag/Service

1. Test device

3. Zero tuning

ADMAG SE :
Zero tuning
Enable
Inhibit
Enable

DEL

DEL

ESC

ENTER

Select “Enable” and press ENTER (F4) and SEND (F2).

Press OK (F4).

Wait for about 20 seconds.

Press OK (F4).

This parameter is only for execution.

Call up “Zero tuning” display.

1. Test device

2. Auto zero

ADMAG SE :
About to enter auto
zero

DEL

DEL

ABORT

OK

1

F4

2

ADMAG SE :
Auto zero passed

HELP

SEND

ABORT

OK

3

(OK)

F2

(SEND)

F4

(ENTER)

F4

(OK)

Call up “Auto zero” display.

ADMAG SE :
waiting for auto zero
to complete

DEL

DEL

ABORT

ENTER

1. Zero adjustment should be done only when the fluid velocity is completely zero by

closing the valve.

2. Do not turn off the power to the ADMAG SE immediately after zero adjustment.

Powering off within 20 seconds after zero adjustment will return adjustment to the
previous setting.

IMPORTANT