11. PARAMETER LIST

Tl set value

0 to 999999

0

W

Sets forward direction totalization preset value.

E05

Total set

(00) Inhibit

Inhibit

W

Inhibits or enables forward direction totalization and reverse

E04

(01) Enable

direction totalization from being preset and reset, respectively.

Total switch

0 to 999999

0

W

Sets switching level when using status output function as

E06

totalization switch function.

Reverse total

0 to 999999

0

W

Displays and resets reverse direction totalizaed flow rate.

A31

Diff. total

–999999 to 999999

0

R

Displays differential totalized flow rate between forward and

A32

reverse.
(Diff. total = Totl – Reverse total)

Analog low cut

0 to 10%

0

W

Sets low cut width of current output.

G01

Analog low lmt

–20 to 100%

–20

W

Sets low limit of current output.

G02

Analog hi lmt

0 to 120%

120

W

Sets high limit of current output.

G03

PV AO Alrm typ

Hi

R

Displays burn out direction of current output.

Analog alarm out

(00) 2.4 mA or less

21.6 mA

W

Selects current output during alarm occurrentce.

b11

(01) 4.0 mA

or more

(02) Hold
(03) 21.6 mA or more

Poll addr

0 to 15

0

W

Sets polling address when multidrop mode.

Num req preams

5

5

R

Displays number of request preambles.

Burst mode

Off

Off

W

Selects the mode for the burst mode functionality.

On

Burst option

PV

PV

W

Selects sending items (instantaneous flow rate, output in %,

% range/current

totalization value and/or current output) when burst mode.

Process vars/crnt

Output function

(00) Pulse out

Pulse out

W

Selects status output functions.

b10

(01) Alarm out
(02) Bi direction
(03) Auto 2 ranges
(04) Low alarm
(05) Total switch

Reverse span

0.0001 to 30000

1.0000

W

Sets span for reverse direction.

b30

Bi direction hys

0 to 10%

2

W

Sets hysteresis when change in flow direction.

b31

Forward span2

0.0001 to 30000

1.0000

W

Sets the second span for forward direction.

b33

Auto range hys

0 to 15%

10

W

Sets hysteresis when change in Auto 2 ranges.

b34

Low alarm

–10 to 110%

–10

W

Sets low limit alarm.

b36

Low alarm hys

0 to 10%

5

W

Sets hysteresis for low limit alarm.

b37

Pulse unit

(00) n UNIT/P

PULSE/s

W

Selects pulse rate unit.

F01

(01) u UNIT/P

(“UNIT” in the selection on the left represents the “Base vol

(02) m UNIT/P

unit” selected.)

(03) UNIT/P
(04) k UNIT/P
(05) M UNIT/P
(06) PULSE/s

Name

Data Range, Units

Default

Value

R/W

Description