Product Service 1-800-522-7658

10

Quickstart


How do I recalibrate the

blower speed

filter life gauge

airflow sensor reading

of the Axiom?

1.

Press the

Menu

button.

2.

Press the

button until the

Services

option is highlighted (it will

turn yellow).

3.

Press

OK/Mute

to enter the Services menu screen.



4.

Press the

button until the

Calibration

option is highlighted (it

will turn white).

5.

Press

OK/Mute

to start the calibration procedure.





6.

You will see a

warning screen

alerting you that you are about to

alter the BSC’s settings.

7.

Press

OK/Mute

to input the password.




8.

When requested for the password press

Light

,

UV light

,

Timer

,

Timer

then

OK/Mute

.





9.

If the password is properly entered, the first certification screen

will display the supply blower motor’s PWM setting. Press the

or

button to increase or decrease the blower speed as

needed. When the speed is set properly, press the

OK/Mute

button to lock the new blower speed setting, and continue to the

filter life screen.

MAIN MENU

1. MyLogic

TM

2. Set t ings

3. Service

 

or

t o choose,

OK t o select , MENU t o escape

or

t o choose,

OK t o select , MENU t o ret urn

Current Settings

Restore Default

Diagnostics

Configuration

Calibration

 

Service

ATTENTION!

The following screens are used

to alter the operation of the

Logic. They should only be used

by a qualified certifier as part

of the certification process.

Calibrat ion

OK t o proceed, MENU to return

PASSWORD:

Calibrat ion

The ECM set t ing is: 80

The current RPM is: 1050

The average RPM is: 1020

Calibration

Supply Fan Speed

or

t o change mot or speed,

OK t o select , MENU t o escape

 