Labconco Type C1 Biosafety Cabinet Models 30468 User Manual
Page 22

Product Service 1-800-522-7658
22
QuickChart – Axiom Type C1 BSCs – Metric Units
Model
30441
30448
30461
30468
Type
C1
C1
C1
C1
Cabinet Size (in m)
1.22
1.22
1.83
1.83
Sash Opening (mm)
254
203
254
203
Starting Serial #
1503_
1503_
1503_
1503_
Downflow Data
Nominal Avg. Downflow (FPM)
65+/-5
65+/-5
55+/-5
55+/-5
grid # of points (rows x columns)
24 (3x8)
24 (3x8)
36(3x12)
36(3x12)
Grid distance from back & sides (mm)
152
152
152
152
Distance between rows (mm)
141
146
141
146
Distance between columns (mm)
132
132
140
140
Inflow Data
Nominal Average Inflow (m/s)
.533+/-.03
.533+/-.03
.533+/-.03
.533+/-.03
Sash Open Area (Sq. m)
.313
.250
5.03
4.03
Nominal Avg. Exhaust Vol. (m3/s)
354
283
.468
.374
Avg. Exhaust Vol. Range (m3/s)
337-371
269-296
503-553
403-443
Secondary Inflow Data
Sash Opening Template
1
4+
4+
6+
6+
Sensor distance (mm)
2
83
83
102
102
# of Test points
8
8
12
12
Test point location
*
*
*
*
Nominal Avg. Inflow Vel. (m/s)
1.80
1.44
1.47
1.17
Avg. Inflow Vel. Range (m/s)
1.71-1.88
1.37-1.50
1.40-1.53
1.12-1.23
Correction Factor (CF)
.0929
.0929
.1700
.1700
Average Inflow Volume Range
(AIV) (Avg. velocity x CF)
.159-.175
.127-.140
.238-.261
.190-.209
Sash Open Area (Sq. m)
.313
.250
.468
.374
Inflow Velocity Range
(AIV / Sash open area)
.533+/-.03
.533+/-.03
.533+/-.03
.533+/-.03
Supply HEPA Data
Air Displacement (m3/s)
.250
.250
.311
.311
Laskin Nozzles needed
1
1
1
1
Theoretical aerosol conc. (ug/l)
3
25
25
20
20
Width x Depth x Height (in.)
4
48x18x3.06
48x18x3.06
72x18x3.06
72x18x3.06
Performance (CFM)
445
445
665
665
Performance (Pressure in. H
2
O)
<.45”
<.45”
<.42”
<.42”
Labconco P/N
3838401
3838401
3838403
3838403
Exhaust HEPA Data
Air Displacement (m3/s)
.167
.134
.250
.200
Laskin Nozzles needed
1
1
1
1
Theoretical aerosol conc. (ug/l)
3
38
48
26
32
Width x Depth x Height (in.)
4
26x18x5.88
26x18x5.88
48x18x5.88
48x18x5.88
Performance (CFM)
283
283
423
423
Performance (Pressure in. H
2
O)
<0.3
<0.3
<0.3
<0.3
Labconco P/N
3838501
3838501
3838503
3838503
Exhaust Volume Data
DIM Exhaust Volume (m3/s)
5
.189
.152
.269
.219
Traverse Exhaust Volume (m,3/s)
6
.227
.183
.323
.262
Differential Pressure (Pascal)
7
75
75
75
75
IRV Inflow Range (m/s)
.010-.047
.010-.047
.010-.047
.010-.047
Nominal IRV Slot Velocity (m/s)
1.22-1.32
1.22-1.32
1.22-1.32
1.22-1.32
Nominal IRV Slot Volume (m3/s)
.235
.235
.235
.235
Supply Motor/Blower Data
Motor HP
1/3
1/3
1/3
1/3
Exhaust Motor/Blower Data
Motor HP
1/3
1/3
1/3
1/3
Fluorescent Lamp Data
Fluorescent Lamps (2 each)
F32T8 TL741
F32T8 TL741
F40T8 SP41
F40T8 SP41
Fluor. Lamp Labconco Part #
9721900
9721900
9721903
9721903
Fluor. Ballast Labconco Part #
3838100
3838100
3838100
3838100