QuickChart – Axiom Type C1 BSCs – Metric Units

Model

30441

30448

30461

30468

Type

C1

C1

C1

C1

Cabinet Size (in m)

1.22

1.22

1.83

1.83

Sash Opening (mm)

254

203

254

203

Starting Serial #

1503_

1503_

1503_

1503_

Downflow Data

Nominal Avg. Downflow (FPM)

65+/-5

65+/-5

55+/-5

55+/-5

grid # of points (rows x columns)

24 (3x8)

24 (3x8)

36(3x12)

36(3x12)

Grid distance from back & sides (mm)

152

152

152

152

Distance between rows (mm)

141

146

141

146

Distance between columns (mm)

132

132

140

140

Inflow Data

Nominal Average Inflow (m/s)

.533+/-.03

.533+/-.03

.533+/-.03

.533+/-.03

Sash Open Area (Sq. m)

.313

.250

5.03

4.03

Nominal Avg. Exhaust Vol. (m3/s)

354

283

.468

.374

Avg. Exhaust Vol. Range (m3/s)

337-371

269-296

503-553

403-443

Secondary Inflow Data

Sash Opening Template

1

4+

4+

6+

6+

Sensor distance (mm)

2

83

83

102

102

# of Test points

8

8

12

12

Test point location

*

*

*

*

Nominal Avg. Inflow Vel. (m/s)

1.80

1.44

1.47

1.17

Avg. Inflow Vel. Range (m/s)

1.71-1.88

1.37-1.50

1.40-1.53

1.12-1.23

Correction Factor (CF)

.0929

.0929

.1700

.1700

Average Inflow Volume Range

(AIV) (Avg. velocity x CF)

.159-.175

.127-.140

.238-.261

.190-.209

Sash Open Area (Sq. m)

.313

.250

.468

.374

Inflow Velocity Range

(AIV / Sash open area)

.533+/-.03

.533+/-.03

.533+/-.03

.533+/-.03

Supply HEPA Data

Air Displacement (m3/s)

.250

.250

.311

.311

Laskin Nozzles needed

1

1

1

1

Theoretical aerosol conc. (ug/l)

3

25

25

20

20

Width x Depth x Height (in.)

4

48x18x3.06

48x18x3.06

72x18x3.06

72x18x3.06

Performance (CFM)

445

445

665

665

Performance (Pressure in. H

2

O)

<.45”

<.45”

<.42”

<.42”

Labconco P/N

3838401

3838401

3838403

3838403

Exhaust HEPA Data

Air Displacement (m3/s)

.167

.134

.250

.200

Laskin Nozzles needed

1

1

1

1

Theoretical aerosol conc. (ug/l)

3

38

48

26

32

Width x Depth x Height (in.)

4

26x18x5.88

26x18x5.88

48x18x5.88

48x18x5.88

Performance (CFM)

283

283

423

423

Performance (Pressure in. H

2

O)

<0.3

<0.3

<0.3

<0.3

Labconco P/N

3838501

3838501

3838503

3838503

Exhaust Volume Data

DIM Exhaust Volume (m3/s)

5

.189

.152

.269

.219

Traverse Exhaust Volume (m,3/s)

6

.227

.183

.323

.262

Differential Pressure (Pascal)

7

75

75

75

75

IRV Inflow Range (m/s)

.010-.047

.010-.047

.010-.047

.010-.047

Nominal IRV Slot Velocity (m/s)

1.22-1.32

1.22-1.32

1.22-1.32

1.22-1.32

Nominal IRV Slot Volume (m3/s)

.235

.235

.235

.235

Supply Motor/Blower Data

Motor HP

1/3

1/3

1/3

1/3

Exhaust Motor/Blower Data

Motor HP

1/3

1/3

1/3

1/3

Fluorescent Lamp Data

Fluorescent Lamps (2 each)

F32T8 TL741

F32T8 TL741

F40T8 SP41

F40T8 SP41

Fluor. Lamp Labconco Part #

9721900

9721900

9721903

9721903

Fluor. Ballast Labconco Part #

3838100

3838100

3838100

3838100