Configuring the Axiom

In the first MyLogic screen, use the

and

buttons on the touchpad to highlight

configure my

Logic for use

-it will turn white when selected. Press

OK/MUTE

to enter the first configuration screen:

The first screen gives you the option of activating

the Smart-Start option for the blower; if you want

the cabinet blower to start every time you raise the

sash, select

start

, and then press

OK/Mute

. If

not

start

is selected, then the blower must be manually

started from the keypad. When

OK/MUTE

is

pressed, the next configuration screen will appear.

The next screen gives you the option of activating

the Smart-Start option for the fluorescent light; if

you want the lights to turn on every time you raise

the sash, select

turn on

, and then press

OK/Mute

.

If

stay off

is selected, then the lights must be

manually turned on from the keypad. When

OK/MUTE

is pressed, the next configuration screen

will appear.

If you want the cabinet blower to run slowly,

maintaining reduced airflows every time you close

the sash, select

go into NightSmart mode

and then

press

OK/Mute

. If

stop

is selected, then the

blower will stop when the sash is closed. When

OK/MUTE

is pressed, the next configuration screen

will appear.


If your Axiom is configured for a UV light, then you

will see the next two screens; if you want the UV

lamp to turn on every time you close the sash,

select

turn on

, and then press

OK/Mute

. If

stay

off

is selected, then the UV light will not turn on

when the sash is closed. When

OK/MUTE

is

pressed, the final configuration screen will appear.

Advertising