Prerequisites, Space requirements, Clearance – Labconco Type C1 Biosafety Cabinet Models 30468 User Manual

Page 37

Advertising
background image

Product Service 1-800-522-7658

37

Section Two –
Installation Requirements

This section offers an overview of site requirements and how to prepare the

unit for operation.

Prerequisites

Before you install the Axiom, you need to prepare the site for installation.

Examine the location where you intend to install the cabinet. You must be

certain that the area is level and of solid construction. In addition, a

dedicated source of electrical power must be located near the installation

site.
Carefully read this chapter to learn:

Location requirements.

Electrical power requirements.

Exhaust requirements.

Service utility requirements.

Space requirements.

Refer to Section Nine: Miscellaneous, for complete biosafety cabinet

electrical and environmental conditions, specifications and requirements

.

Space Requirements

The overall dimensions for the 4-foot Axiom are 64.2 inches (163 cm) high, 32

inches (81 cm) deep, and 54.2 inches (138 cm) wide. The overall dimensions

for the 6-foot Axiom are 64.2 inches (163 cm) high, 32.6 inches (83 cm) deep,

and 78.2 inches (199 cm) wide.

Complete dimensions for the Axiom biosafety cabinets are shown in

Section

Nine: Miscellaneous

.

Clearance

A minimum clearance of at least 4 inches (100 mm) is suggested on the top

and 6 inches (150mm) on both sides of the cabinet for service.