Product Service 1-800-522-7658

134

5.

Using the screws, flat washers and locking nuts provided, secure the UV ballast to

the top of the electronics module.

6.

Wire the UV ballast into the wiring harness, following the wiring diagrams located in

Figure 2a-c.

7.

Reattach the electronics module to the top of the cabinet before reconnecting the

electrical power.

8.

Raise the sash, and install the UV lamp by rotating the lamp pins into the sockets.

9.

For all Logic Models:

Lower the sash to fully closed position, plug in the unit, and turn the system

power switch to the "UV" position. The UV lamp will illuminate. Refer to the

user’s manual for further information regarding the proper operation and

maintenance of the UV light.

For All Axiom models:

Lower the sash to fully closed position, plug in the unit, and do the following:


1.

Press the

Menu

button.

2.

Press the

button until the

Services

option is highlighted (it will

turn yellow).

3.

Press

OK/Mute

to enter the Services menu screen.





4.

Press the

button until the

Configuration

option is highlighted

(it will turn white).

5.

Press

OK/Mute

to alter the configuration.



6.

You will see a

warning screen

alerting you that you are about to

alter the BSC’s settings.

7.

Press

OK/Mute

to input the password.






8.

When requested for the password press

Light

,

UV light

,

Timer

,

Timer

then

OK/Mute

.


MAIN MENU

1. MyLogic

TM

2. Set t ings

3. Service

 

or

t o choose,

OK t o select , MENU t o escape

or

t o choose,

OK t o select , MENU t o ret urn

Current Settings

Restore Default

Diagnostics

Calibration

Configuration

 

Service

Conf igurat ion

PASSWORD:

ATTENTION!

The following screens are used

to alter the operation of the

Logic. They should only be used

by a qualified certifier as part

of the certification process.

Conf igurat ion

OK t o proceed, MENU to return