Section 5
Certification Procedures for the Axiom
This section will outline test procedures to validate the performance of the Axiom.
Note: These are the only Labconco-approved test procedures for validating
the performance of the Axiom Biosafety Cabinet. Other test procedures may
yield different or inaccurate test results.
NOTE: In order to properly certify the Axiom, you must obtain an Axiom
Certifier Kit, Labconco P/N 3858400. Qualified certifiers can obtain this kit by
contacting Labconco Customer Service Department at 800-821-5525.
Note: The ECM motor is programmed by Labconco to deliver a constant
volume of air throughout the life of the HEPA filters. Any changes to the
blower speed control should be small (less than 5 units).
Accessing the Certifier’s Menu
Note: You will need to access this menu to do any of the following:
Adjust the blower speed
To reset the filter gauge
To calibrate the optional airflow sensor
Adjusting the Downflow, Inflow and Filter Life Gauges
1.
Press the
Menu
button.
2.
Press the
▼
button until the
Services
option is highlighted (it will
turn yellow).
3.
Press
OK/Mute
to enter the Services menu screen.
4.
Press the
▼
button until the
Calibration
option is highlighted (it
will turn white).
5.
Press
OK/Mute
to start the calibration procedure.
6.
You will see a
warning screen
alerting you that you are about to
alter the BSC’s settings.
7.
Press
OK/Mute
to input the password.
8.
When requested for the password press
Light
,
UV light
,
MAIN MENU
1. MyLogic
TM
2. Set t ings
3. Service
or
t o choose,
OK t o select , MENU t o escape
or
t o choose,
OK t o select , MENU t o ret urn
Current Settings
Restore Default
Diagnostics
Configuration
Calibration
Service
ATTENTION!
The following screens are used
to alter the operation of the
Logic. They should only be used
by a qualified certifier as part
of the certification process.
Calibrat ion
OK t o proceed, MENU to return
PASSWORD:
Calibrat ion