Section 5

Certification Procedures for the Axiom

This section will outline test procedures to validate the performance of the Axiom.

Note: These are the only Labconco-approved test procedures for validating

the performance of the Axiom Biosafety Cabinet. Other test procedures may

yield different or inaccurate test results.
NOTE: In order to properly certify the Axiom, you must obtain an Axiom

Certifier Kit, Labconco P/N 3858400. Qualified certifiers can obtain this kit by

contacting Labconco Customer Service Department at 800-821-5525.
Note: The ECM motor is programmed by Labconco to deliver a constant

volume of air throughout the life of the HEPA filters. Any changes to the

blower speed control should be small (less than 5 units).

Accessing the Certifier’s Menu

Note: You will need to access this menu to do any of the following:

Adjust the blower speed

To reset the filter gauge

To calibrate the optional airflow sensor

Adjusting the Downflow, Inflow and Filter Life Gauges

1.

Press the

Menu

button.

2.

Press the

button until the

Services

option is highlighted (it will

turn yellow).

3.

Press

OK/Mute

to enter the Services menu screen.



4.

Press the

button until the

Calibration

option is highlighted (it

will turn white).

5.

Press

OK/Mute

to start the calibration procedure.





6.

You will see a

warning screen

alerting you that you are about to

alter the BSC’s settings.

7.

Press

OK/Mute

to input the password.



8.

When requested for the password press

Light

,

UV light

,

MAIN MENU

1. MyLogic

TM

2. Set t ings

3. Service

 

or

t o choose,

OK t o select , MENU t o escape

or

t o choose,

OK t o select , MENU t o ret urn

Current Settings

Restore Default

Diagnostics

Configuration

Calibration

 

Service

ATTENTION!

The following screens are used

to alter the operation of the

Logic. They should only be used

by a qualified certifier as part

of the certification process.

Calibrat ion

OK t o proceed, MENU to return

PASSWORD:

Calibrat ion