Service Line Requirements

All utility service lines should be ¼ inch O.D., brass, copper, or stainless

steel, and equipped with an easily accessible shut-off valve. If the service

line pressure exceeds 40 PSI, it must be equipped with a pressure regulator

to reduce the line pressure.
Note: The use of flammable gases or solvents should be avoided in the

biosafety cabinet. Open flame in the cabinet will disrupt the laminar airflow

in the cabinet and may damage the HEPA filters. Flammable gases or

solvents may reach explosive concentrations in the cabinet or ductwork. If

you feel that the procedure requires the use of an open flame or flammable

materials, contact the institution’s safety office.
The use of air or gases under high pressure should be avoided as they may

seriously disrupt the airflow patterns in the cabinet.