Inflow Velocity Calculation - Primary Method

Note: The dress panel MUST be installed on the cabinet to obtain an accurate inflow

value.
The average inflow velocity for Axiom models is 105 ±5 FPM. It should be determined

by converting the inflow volume to the average inflow velocity. The inflow volume is

measured directly by sealing a direct inflow measuring flow meter to the face of the

unit. When corrected for local conditions, dividing the inflow volume by the opening area

will yield the average inflow velocity.

Table 5-4

Inflow Test Specifications – 8 inch Sash Opening

Width (feet)

4

6

Nominal Average Inflow

(FPM)

105+/-5

105+/-5

Sash Open Area (Sq. Ft)

2.69

4.03

Nominal Avg. Exhaust Vol.

(CFM)

283

423

Avg. Exhaust Vol. Range

(CFM)

269-296

403-443

Inflow Test Specifications – 10 inch Sash Opening

Width (feet)

4

6

Nominal Average Inflow

(FPM)

105+/-5

105+/-5

Sash Open Area (Sq. Ft)

3.37

5.03

Nominal Avg. Exhaust Vol.

(CFM)

354

529

Avg. Exhaust Vol. Range

(CFM)

337-371

503-553

Acceptance
Average inflow shall be meet the specified value.