129


6.

Note the orientation of the sensor

harness plug and the sensor; it is a

keyed connection-and can only be

connected in the correct orientation.

Plug the sensor harness into the

sensor, and the other end into the

multi pin connector on the Axiom’s

electronics module.


7.

Reconnect the Logic to electrical service. The unit is now ready for the calibration

of the air flow sensor.

Note: You will need to access the certifiers menu to calibrate the airflow

sensor.

Note: To properly calibrate the air flow sensor, a qualified certifier must

verify the inflow and downflow values.

With the Axiom operating:

8.

Press the

Menu

button.

9.

Press the

button until the

Services

option is highlighted (it will

turn yellow).

10.

Press

OK/Mute

to enter the Services menu screen.



11.

Press the

button until the

Calibration

option is highlighted (it

will turn white).

12.

Press

OK/Mute

to start the calibration procedure.





13.

You will see a

warning screen

alerting you that you are about to

alter the BSC’s settings.

14.

Press

OK/Mute

to input the password.



MAIN MENU

1. MyLogic

TM

2. Set t ings

3. Service

 

or

t o choose,

OK t o select , MENU t o escape

or

t o choose,

OK t o select , MENU t o ret urn

Current Settings

Restore Default

Diagnostics

Configuration

Calibration

 

Service

ATTENTION!

The following screens are used

to alter the operation of the

Logic. They should only be used

by a qualified certifier as part

of the certification process.

Calibrat ion

OK t o proceed, MENU to return