The exhaust system must be capable of moving the following volumes of
exhaust air at the negative pressures listed. The Airflow Volumes are the
values recorded via a direct measurement using a flow hood at the cabinet.
The Concurrent Balance Values are measured in the exhaust duct via
traverse methodology, and will always be higher due to differences in volume
measurement methodologies.


Axiom Type C1

Airflows

Airflow Volume

Concurrent

Balance Value

Recommended

Duct Vacuum

1

ft

3

/min

m

3

/sec

ft

3

/min

m

3

/sec

WC

Pa

4-foot, 8" Sash

323

0.15

387

0.18

0.30

75

4-foot, 10" Sash

400

0.19

480

0.23

0.30

75

6-foot, 8" Sash

463

0.22

556

0.26

0.30

75

6-foot, 10" Sash

570

0.27

684

0.32

0.30

75


1 – Unlike Type B cabinets, the recommended vacuum will remain constant
throughout the life of the Exhaust HEPA filter. Duct vacuums below 0.3 WC
(37 Pa), or above 0.5 WC (125 PA) may result in erratic operation.