Testing a circuit breaker, Replacing a circuit breaker – Labconco Type C1 Biosafety Cabinet Models 30468 User Manual

Page 118

Advertising
background image

Product Service 1-800-522-7658

118

Circuit Breaker Specifications, 230 VAC

Cabinet Width (feet)

4

6

Motor/Light Breaker Rating,

Amps

6

6

Motor/Light Breaker, Labconco

Part #

1327203

1327203

Motor/Light Breaker,

Mechanical Products Part #

1600-062-

060-007

1600-062-

060-007

Outlet Breaker Rating, Amps

3

3

Outlet Breaker, Labconco

Part #

1327200

1327200

Outlet Breaker, Mechanical

Products Part #

1600-062-

030-007

1600-062-

030-007


Testing a circuit breaker

Note: If the breaker is tripped (white button is extended), press the reset button

of the breaker. If it does not reset smoothly, the breaker has failed and must be

replaced.

1.

With the breaker reset, unplug the Axiom, either at the wall, or at the power

cord inlet at the back of the electronics module, as shown in Figure 7-2.

2.

Remove the three 7/16-inch nuts that secure the module.

3.

Tip the module on its side to access the breakers.

4.

Using a continuity meter, check the continuity (resistance) across the

terminals of the breaker being tested.

5.

The resistance across the breaker should not exceed the resistance noted

when the meter probes are connected to each other.

Replacing a circuit breaker

1.

Unplug the Axiom, either at the wall, or at the power cord inlet at the back of

the electronics module, as shown in Figure 7-2.

2.

Remove the three 7/16-inch nuts that secure the module.

3.

Tip the module on its side to access the breakers.

4.

Disconnect the wires connected to the breaker terminals.

5.

Using a pair of pliers, loosen the chrome ring nut outside of the reset button.

Remove the ring nut, and pull the breaker out of the electronics module.