4.

If connected to an exhaust

system, secure the Inlet

Relief Valve in an open

position, to keep the cabinet

blowers operating, and then

remove the screws that

secure the front of the

exhaust connection to its

structure. Place the front

panel aside.

5.

The Axiom is now ready for the Exhaust HEPA Filter Leak Test. Set the

photometer sampling valve to “DOWNSTREAM”. Ensure there is a proper
vacuum at the sampling tube.

6.

Open the Laskin nozzle(s).

7.

Ensure the pressure gauge on the

generator is reading 22 +/- 1

PSIG.

8.

Using a rigid 18” long x 3/8” o.d.

tube connected to the

photometer probe pickup, Scan

the downstream side of the filter,

resting the tube in the ridges

located on the top of the opening.

By resting the tubing in

successive ridges, and

alternatively pushing the probe in

on the first ridge, then shifting

the tube to the next ridge, and

then pulling it out, you can

quickly scan the filter surface in

overlapping strokes. By resting

your index finger under the tube

during scanning, you can

accurately control the probe

position and height.

9.

Observe the average

concentration of aerosol
downstream.

Acceptance
Average aerosol penetration shall not exceed 0.01%.