Quickstart – Labconco Type C1 Biosafety Cabinet Models 30468 User Manual

Page 15

Advertising
background image

Product Service 1-800-522-7658

15

Quickstart


How do I run a diagnostic on the sash position switches?


1.

Press the

Menu

button.

2.

Press the

button until the

Services

option is highlighted (it will

turn yellow).

3.

Press

OK/Mute

to enter the Services menu screen.





4.

Press the

button until the

Diagnostic

option is highlighted (it

will turn white).

5.

Press

OK/Mute

to enter the Diagnostic submenu.






6.

Press the

button until the Diagnostic

Sash sensors

option is

highlighted (it will turn white). Press

OK/Mute

.





7.

Lower the sash, and note the status line – it should properly

report the sash as open too high, open, nearly closed or closed

when the sash is in the corresponding position.* Press

OK/Mute

to exit the diagnostic.





MAIN MENU

1. MyLogic

TM

2. Set t ings

3. Service

 

or

t o choose,

OK t o select , MENU t o escape

or

t o choose,

OK t o select , MENU t o ret urn

Current Settings

Restore Default

Calibration

Configuration

Diagnostics

 

Service

Diagnost ics

Test t he alarm by closing t he

inlet valve.

Note the alarm st at us below:

Open

OK t o ret urn

Advertising