Labconco Type C1 Biosafety Cabinet Models 30468 User Manual

Page 106

Advertising
background image

Product Service 1-800-522-7658

106

filter and set it in a safe spot.

7.

Disconnect the exhaust motor harness at the pass through near the top of the

chamber, as shown in Figure 6-17.

Figure 6-17

Exhaust Motor wiring harness

connectors




NOTE: It will ease plenum installation and removal by laying two 2x4 or 4x4

supports under the plenum, and over the filter chamber flange, as shown in Figure

6-18.

Figure 6-18





8.

Loosen the filter clamping bolts until the plenum assembly is released.

The motor/blower/plenum assembly is heavy. Handle with care.


9.

Slide the plenum out of the cabinet on the wooden supports.

10.

Disconnect the exhaust motor wiring harness from the ECM Motor, and install it

on the new motor/blower/plenum assembly.

11.

Install the new assembly into the Axiom.

12.

To complete the replacement, reverse the above procedure.

13.

Plug in the cabinet and recertify it before use.