Labconco Type C1 Biosafety Cabinet Models 30468 User Manual

Page 133

Advertising
background image

Product Service 1-800-522-7658

133

UV Lamp Kit No. 3858500, -01, -02

All Labconco Purifier Axiom Series Biosafety Cabinets are pre-wired for the installation of

a UV lamp kit, consisting of the UV lamp, its ballast, and fasteners. The kit is part

number 3858500 for 115 Volt models, 3858501 for 100 Volt models, and 3858502 for

230 Volt units.
A qualified electrician or certifier should install the UV lamp kit.
Decontaminate all interior work surfaces of the Purifier before beginning the

installation of the UV lamp.

Installation Procedure
1.

Disconnect power to the unit.

2.

Using a 7/16-inch wrench or socket, remove the two acorn nuts that secure the

electronics module to the top of the Purifier.

3.

Tilt the electronics module on its side to access the underside of the module. Secure

the module so that it does not fall and damage the wiring.

4.

Locate the fluorescent lamp ballast in the module. The UV lamp ballast mounts next

to the fluorescent ballast, as shown in Figure 1.


UV Ballast


Figure 1

Advertising