RS-232 Connection Kits

RS-232 Kit 3858310 (9 Pin), 3858311 (25 Pin)

The kit contains the RS-232 PCB (P/N 3402200), the attaching hardware and connecting

cables to install the PCB inside the electrical control box (located on top of the Axiom).

Picture 1 shows an electrical control box with the Axiom RS-232 PCB installed.

Picture 1

WARNING!! DISCONNECT ALL ELECTRICAL POWER FROM THE UNIT BEFORE

BEGINNING THIS PROCEDURE!!!!

To install:

1)

Carefully remove the square “knock-outs” shown in picture 2 under the “RS-232

PORT” label and above the “ALARM CONTACT” label.

Picture 2

2)

Locate the black square connecter labeled as J2 and the white connector labeled

as J3 on the RS-232 PCB, see picture 3. These two connectors show through the

Advertising