Labconco Type C1 Biosafety Cabinet Models 30468 User Manual

Page 54

Advertising
background image

Product Service 1-800-522-7658

54

Disconnecting the Axiom from an Exhaust System Connection to Operate in
Reciculation

a)

Mechanical configuration

1.

Unplug the Axiom. Using an appropriate ladder or platform, remove the

exhaust cover panel on the top left side of the exhaust cover, and install it

on the studs on the right side. Hand tighten the wing nuts to secure it, as

shown in figure 3-1a.

2.

Disconnect the Axiom’s connection collar from the exhaust system, and cap

the building’s exhaust connection in a way that meets all appropriate codes
and regulations. Leave the Axiom’s exhaust collar open to allow for

unrestricted airflow out of the unit.



Exhaust Cover Panel


Figure 3-1a