Figure 5-11


The Sensor Distance, as described in Table 5-5,

using the slots cut in the 6-foot Logic template.

4.00-inch mark

3.25-inch mark

Figure 5-12

The Sensor Distance, as described in Table 5-5,

using a scale.




Table 5-5

Secondary Inflow Test Specifications – 8 Inch Sash Opening

Cabinet Width (feet)

4

6

Sash Opening Template

4+

6+

Sensor distance (inches)

3.25

4

# of Test points

8

12

Test point location

*

*

Nominal Avg. Inflow Vel. (FPM)

283

231

Avg. Inflow Vel. Range (FPM)

249-274

221-222

Correction Factor (CF)

1.08

2

Average Inflow Volume Range (AIV) (Avg. velocity x CF)

269-296

403-443

Sash Open Area (Sq. Ft)

2.69

4.03

Inflow Velocity Range (AIV / Sash open area)

100-110

100-110

Secondary Inflow Test Specifications – 10 Inch Sash Opening

Cabinet Width (feet)

4

6

Sash Opening Template

4+

6+

Sensor distance (inches)

3.25

4

# of Test points

8

12

Test point location

*

*

Nominal Avg. Inflow Vel. (FPM)

354

289

Avg. Inflow Vel. Range (FPM)

312-344

251-276

Correction Factor (CF)

1.08

2

Average Inflow Volume Range (AIV) (Avg. velocity x CF)

337-371

503-553

Sash Open Area (Sq. Ft)

3.37

5.03

Inflow Velocity Range (AIV / Sash open area)

100-110

100-110