Sash Position Switches

Four microswitches are used to indicate the sash position. The switches are the same,

Labconco part # 3832400. A photo of the switches is shown in Figure 6-6.

Figure 6-6

Sash Closure Position Switch


Sash Restricted Opening Switch

8” Sash Operating Position Switch

10” Sash Operating Position Switch

Sash Closure Position Switch

The sash closure position switch signals the display microprocessor that the sash

is open or closed allowing proper operation of the Blowers, fluorescent or UV

lights. It is wired in the normally open configuration. If the switch does not

actuate properly when the sash is within ¼-inch of being fully closed, ensure the

switch follower is properly actuating against the weight. See Figure 6-7.

Sash Restricted Opening Switch

The sash restricted opening switch signals the display microprocessor that the

sash is about to close. When actuated, the blowers are shut off to prevent the

exhaust blower from forming a vacuum when the sash is closed. If the switch

does not actuate properly when the sash is within 2 inches of being fully closed,

ensure the switch follower is properly actuating against the weight. See Figure 6-

7.