Supply motor/blower replacement – Labconco Type C1 Biosafety Cabinet Models 30468 User Manual

Page 103

Advertising
background image

Product Service 1-800-522-7658

103

To Install a Supply HEPA Filter:

1.

Install a new supply HEPA filter by pushing it straight into the cabinet,

ensuring that it is correctly oriented, and that the filter is centered on the

filter shelf.

2.

Use the 1/2-inch sockets or wrenches to sequentially loosen the three

plenum lift bolts located as noted in Figure 6-12. When properly positioned,

the plenum box will be aligned with the supply HEPA filter frame. Continue to

loosen the bolts until the plenum contacts the HEPA filter and has slightly

compressed the filter gasket.


NOTE: The lift bolts should only be tightened enough to ensure a proper seal

of the supply HEPA filter. At maximum tightness, the supply filter's gasket

should be compressed to less than 50% of its original thickness. NEVER

tighten the locking bolts beyond this setting.


Supply Motor/Blower Replacement

Caution: The cabinet must be properly decontaminated before servicing the

motor/blower.

The cabinet blower motor's bearings are sealed and require no lubrication.

DO NOT attempt to lubricate them.
To replace the Motor/Blower:

1.

Unplug the cabinet.

2.

Remove the front dress panel as shown in Figure 6-1.

3.

Loosen all of the Phillips screws on the blower chamber cover. Remove all of the

screws except two screws on the top center of the panel. This will support the

panel.

4.

Remove the last two screws. Gently rock the cover, slide it to the left or right,

and swing it out and away from the unit.

NOTE: During disassembly, the blower plenum cover can rest on the central sash

support, as shown in Figure 6-11, to allow for repositioning of the panel or your grip on

it.

5.

Disconnect both motor connectors from the motor. The 5-pin power connector,

located closest to the blower, is released by pressing the locking ears on either

side of the connector. The 16-pin data connector, located furthest from the

blower, is released by pressing the single locking ear located on the inside center

surface of the connector body. The connectors are shown in Figure 6-13.