Using the RS-232 Receptacle

The operation of the Axiom Biological Safety Cabinet (BSC) can be monitored using a

computer connected to the RS232 receptacle located on the top electrical box of the

Axiom. The computer monitors the Axiom’s operation, but cannot control the operation

of the Axiom.

Connection for Computer Interface

Check your computer to see which type of RS232 serial port is provided. Use one of the

two connecting cables listed below:

1.

Computers with a 25-pin D-sub male serial connector should use Connect Cable,

Labconco part number 7537801, to connect a computer to the Axiom.

2.

Computers with a 9-pin D-sub male serial connector should use Connect Cable,

Labconco part number 7537800, to connect a computer to the Axiom.


* Note – If your computer does not have an RS-232 serial port a USB to RS232 adapter

can be purchased at most computer stores, discount stores, etc. to allow the Axiom

cable to connect to a USB port on your computer.

The purpose of the RS232 interface is to send data to a data logging computer to

monitor the state and activity of the Axiom. This data is half duplex data. The data

properties are as follows:

1.

Data Rate 2400 Baud

2.

8 bit word length

3.

1 Start bit, 1 Stop bit

4.

No parity is transmitted

5.

Standard ASCII character set

The time between data transmissions may be varied by the user to occur at 1, 10, 30 or

60 second intervals. Using the Axiom control buttons and the menus on the color

display, set the data transmission time by pressing the “MENU” button, then select

“SETTINGS”, then select “RS-232 OUTPUT”. From this display screen the data

transmission intervals can be selected.
The format of the transmitted message with typical values is as follows: