Sash Cable Retainer

The sash cable retainers are located at the top of each sash pulley, as shown in

Figure 6-9. If the retainer needs adjustment, Use a 3/8-inch wrench to loosen

the bolt securing the retainer and adjust it until it there is approximately 1/16”

gap between the retainer and the pulley. Retighten the bolt.

Figure 6-9


HEPA Filter Replacement

CAUTION: The cabinet must be properly decontaminated before servicing the
exhaust HEPA filter. Following replacement of either HEPA filter, the cabinet

must be recertified.

CAUTION: HEPA filters can be awkward to handle and heavy. Use appropriate

lifting techniques to remove and handle it. Use two people to remove the

larger filters if possible.

1.

Unplug the cabinet.

2.

Remove the front dress panel as shown in Figure 6-1.

To Remove the Exhaust HEPA Filter:

3.

Loosen all of the Phillips screws on the exhaust filter cover.

4.

Loosen the four sealed exhaust filter clamp bolts on the top of the cabinet

hull, as shown in Figure 6-10. DO NOT remove the bolts; only loosen them

until the exhaust HEPA filter is free of the filter frame.

5.

Pull the exhaust HEPA filter straight out of the cabinet.