If the Axiom is Contaminated

(An upstream concentration cannot be established)

If the Axiom has been used with biohazardous, toxic chemicals, or radioisotopes, the

DOP access line cannot be used. The upstream concentration cannot be measured

directly. Testing at Labconco has shown the actual concentration varies from the

calculated value. Table 5-1 shows the actual vs. calculated concentrations for the Axiom

models.

Table 5-1

Supply HEPA Filter Leak Test Specifications

Model

3044-

3046-

Air Displacement (CFM)

520

665

Laskin Nozzles needed

1

1

Theoretical aerosol conc.

(ug/l)

1

26

20

Actual aerosol conc. (ug/l)

1

13

12

1.

The calculated concentration was established by the formula:

Concentration (ug/l) = (# Laskin nozzles @ 20 PSIG x 13,500)/Volume of air displaced

2.

Based on Mineral oil

If the Axiom is Uncontaminated

(Establishing an Upstream Concentration)

The Axiom utilizes an aerosol access tube for sampling the upstream concentration of

aerosol. If after reading these instructions, you have further questions, please contact

Labconco's Product Service Department.

1.

Remove the center work surface by lifting the handles on either wing work

surface. Pull the center work surface straight out of the Axiom. Repeat for the

right wing work surface.

2.

Locate the access tube on right side of the rear plenum wall, passing through the

towel catch. As shown in Figure 5-2.

Figure 5-2