8”- and 10” Sash Operating Position Switches

The sash operating position switches signals the display microprocessor that the

sash is open too far for proper operation, as shown in Figure 6-8. Both switches

are wired in the normally open configuration, and the display microprocessor

only responds to the appropriate switch selected during configuration. If the

switch does not actuate properly when the sash is within ¼-inch of its operating

position, ensure the switch follower is properly actuating against the weight.

Figure 6-7

Sash Closure Position Switch properly
actuated against the sash weight

Figure 6-8

Sash Operating Position Switch

properly actuated against the sash

weight