When the Axiom’s main power switch is on, the data will be transmitted and

updated at the time intervals selected by the user.

Axiom Relay Contact Outputs:

When equipped with the optional RS-232 contact kit board, Labconco Axiom Biosafety

Cabinet (BSC) models are provided with two (2) dry relay contacts to allow connection

to an external monitoring system, or to switch other air handling units “on” when the

Labconco unit is turned on.

With the optional RS-232 board installed, the Labconco Axiom BSC provides two

contacts: One to indicate the internal Labconco blower is on, (so that external blowers

may also be energized) and the other to indicate airflow errors. These contacts

provided will “short circuit” when the blower is turned on, or when there is an airflow

error. This means that the Labconco contacts are isolated from the Labconco

equipment, and the user must provide the interface circuitry based on the requirements

of the connected equipment.

The dry contacts are accessed at connector J3 of the Axiom Control Box (Labconco

harness P/N 3432300 is available for easy connection). This J3 connector is located on

the electrical control box on top of the BSC, above the “ALARM CONTACT” label as

shown in the picture below: