Alarm Screens

Sash is too high
The sash is open too far for safe operation.



Airflow Alerts
The airflow patterns in the cabinet have changed, resulting

in a sudden change in either motor speed. This is most likely

due to a blockage of the grille or the exhaust filter outlet. It

may also be caused by removal of the work surface(s) while

the cabinet is in operation.

Exhaust Fan Alarm
There has been a failure of the exhaust fan, or its

communication to the cabinet’s control board. DO NOT USE

THE CABINET UNTIL THE PROBLEM HAS BEEN

CORRECTED.

Exhaust System Alarm-Active Protection Mode
When connected to an exhaust system, and the flow drops

below acceptable limits, this screen will appear. When the

timer reaches 0, the supply blower will shut off, and the

exhaust fan will continue to operate for 10 more seconds,

protecting the operator.

Exhaust System Alarm
After the countdown reaches 0, the exhaust fan will shut off

and this screen will be displayed. If this alarm is displayed,

the Axiom blower will need to be turned off, and then back

on again to reset the alarm. Frequent Exhaust System

Alarms indicate unstable or insufficient exhaust flow, and

must be corrected.

System Error
The motors and display circuit board are not communicating

properly. DO NOT USE THE CABINET UNTIL THE PROBLEM

HAS BEEN CORRECTED.