Optional Exhaust Connection Requirements

Certain applications such as working with odorous products or volatile toxic materials

will require the connection of the biosafety cabinet to an exhaust system.
Note: The Type C1 exhaust connection includes an integral inlet relief valve, which

functions as a one-way, or check valve. During operation, the exhaust system draws all

of the cabinet’s exhaust air, plus a volume of room air through the relief valve into the

exhaust duct. The inlet relief valve functions as a “shock absorber” allowing the cabinet

to maintain optimal inflow during changes in room air pressure.
Note: Because the Type C1 cabinet has an integral exhaust fan, the vacuum

requirements of the exhaust system will be much lower than existing Type B units.

Please see the volume and vacuum requirements in Table 2-1, in

If you intend to

connect the biosafety cabinet to an exhaust system:

in

Chapter 2: Prerequisites

Note:

If the research involves the use of toxic compounds or volatile materials, contact

the facility’s safety officer or Labconco to ensure that the biosafety cabinet and its

exhaust system are compatible with the materials you will be working with.